Table of Contents Table of Contents
Previous Page  92 / 146 Next Page
Information
Show Menu
Previous Page 92 / 146 Next Page
Page Background

Purpose:

The Ligurian Medical Physicist working group on Mammo-

graphic Screening analyzed the results of two quality controls, homogeneous

in all centers, in order to provide to the regional network HTA a compre-

hensive indicator as clinical parameter within the obsolescence evaluation

of a system.

Methods and Materials:

The average glandular dose (AGD) and the thresh-

old contrast visibility were chosen as quality indicators. The results of these

tests are represented by a curve that must remain below the curve of ac-

ceptable level and, if possible, also below the curve of the achievable

level.

In order to obtain two indicators, dose and contrast one, the average dis-

tances between the measured and acceptable values has been calculated

for both the physical quantities and then normalized to acceptable values.

Then dose (DI) and contrast indicators (CI) DI and CI have been linked into

an a-dimensional index.

Results:

All the mammography systems deliver an AGD below the accept-

able levels. Depending on DI definition, the higher the value the lower is

the dose compared to the acceptable value. It was noted that the CR devices

deliver an AGD higher than DR ones.

CI showed that CR equipment has a contrast-detail curve worse than the

DR ones. For some mammography (due to some points out of tolerance)

the CI was even below the acceptable level.

Conclusions:

These two tests are not sufficient to comprehensively eval-

uate the equipment, but are indicative of the dose and image quality and

they were uniformly implementable in all the centers.

The overall assessment parameter that was provided to the Ligurian network

HTA will be part of a set of indexes, such as technical, functional and eco-

nomical obsolescence, which will together give advice relating to any

mammography equipment obsolescence.

This benchmark could allow the Ligurian network HTA to suggest in the

future the purchase of equipment with CR technology and to assess which

device prioritizes for upgrades or replacements.

http://dx.doi.org/10.1016/j.ejmp.2016.01.298

B.294

MULTI HOSPITAL EXPERIENCES OF RADIATION DOSE INDEX

MONITORING

S. Pini

* , a ,

S. Mazzocchi

a ,

A. Bruschi

b ,

A. Ghirelli

a ,

E. Rigacci

a ,

G. Zatell

i a .

a

Azienda Sanitaria Di Firenze, Firenze, Italy;

b

Universita’ Degli Studi Di Firenze,

Firenze, Italy

Introduction:

Radiation Dose Index Monitoring (RDIM) software is an im-

portant tool to optimize radiological procedures, however its implementation

need a thorough acceptance test.

Methods and Materials:

In our multi hospital reality (ASF) we begun in

2012 to monitor dose index data from 7 CTs and 2 angiographic units using

CareAnalytics (CA) a non-medical, free-of-charge tool by Siemens. CA pro-

cesses DICOM Radiation Dose Structured Reports (RDSR) stored in PACS.

Moreover in context of an AIFM project, Radiation Dose Monitor (RDM) by

Medsquare has been tested since Gen 2015. RDM is a web based software

solution that collects, controls and analyzes radiation doses delivered to

patients during medical imaging examinations, it supports either modal-

ity integration, or RIS and Pacs integrations and can retrieve dose and

relevant data from RDSR, directly from header DICOM or via MPPS mes-

sages. In ASF RDM server has been connected to the single modalities in

the RIS PACS network: one mammographic installation, 2 CTs and 1

angiographic unit. The aim of this work is to test the consistency of data

processed by either two RDIM software.

Results:

For angiographic procedure we have compared the maximum in-

cident air kerma at the reference point at different times and at different

C arm positions with the dose values obtained by software analysis. For

mammography, data from header DICOM (kV, HVL, breast thickness, AGD)

have been positively compared with RDM data. For CT we have compared

values obtained from console, CA and RDM.

Conclusion:

In our experiences both the software under test seems to be

useful tools to monitor radiation dose index. RDM allows wider applica-

tions and custom made options.

http://dx.doi.org/10.1016/j.ejmp.2016.01.299

B.295

THREE YEARS OF EXPERIENCE WITH DOSE REDUCTION BY MEANS OF

ITERATIVE RECONSTRUCTION ALGORITHM AVAILABLE ON CT SIEMENS

SOMATOM DEFINITION FLASH

M. Poli

*

, C. Bracco, S. Bresciani, A. Maggio, A. Di Dia, A. Miranti,

M. Stasi.

Medical Physics, Candiolo Cancer Institute, FPO-IRCCS, Candiolo (TO),

Italy

Introduction:

Patients in a radiological unit undergo multiple CT scans, re-

sulting in high cumulative radiation exposures. Three years of experience

with dose reduction by means of CT Somatom Definition Flash is here pre-

sented. In particular, the implementation of SAFIRE low-dose CT protocols

and the results in terms of dose (D) are shown. New protocols were first

proposed by the physicist after a phantom study, then the selected pro-

tocols were clinically implemented only after a clinical validation with which

the physician confirmed no loose of diagnostic quality image.

Material and Methods:

Two acquisition protocols were considered: the

thorax-abdomen (TA) and the head (H). Both protocols were optimized and

the new protocols were clinically implemented in the 2014. The recon-

struction kernel was changed from the iterative I30f strength 2 to the I30f

strength 3 for the TA protocol and from the filtered back projection H31s

to the iterative J40s strength 3 for the H protocol. Retrospectively ana-

lyzed were CTDIvol of 100 patients for each year (2013–15) and for each

protocol in order to estimate the dose reduction gained since the intro-

duction of these new protocols.

Results:

The average CTDIvol relative to the pre-optimization protocol was

(13.7

±

1.8) mGy for the TA and 60.8 mGy for the H protocol. The TA pro-

tocol optimization led to a dose reduction between 2013 and 2014 of the

30%: the average CTDIvol relative to 2014 and 2015 was (9.9

±

1.5) mGy

and (9.7

±

0.6) mGy, respectively. Similarly, we obtained a dose reduction

of 10% for the H protocol, equivalent to an average CTDIvol of 54.7 mGy

in the 2015.

Conclusions:

Protocols optimization by the use of iterative reconstruc-

tion methods, such as SAFIRE, can lead to an important dose reduction.

http://dx.doi.org/10.1016/j.ejmp.2016.01.300

B.296

ITERATIVE METHOD FOR CT SYSTEM: DOSE REDUCTION AND

QUANTITATIVE ANALYSIS OF IMAGE QUALITY IMPROVEMENT

M. Porzio

*

, F. Coloberti, E. Zucchi.

ASL 1 Imperiese, Sanremo, Italy

Introduction:

This work analyzes a commercial iterative method (Adap-

tive Statistical Iterative Reconstruction-ASiR – GE Medical Systems) for which

noise reduction technique may allow for dose reduction without loss of

image noise levels or be applied to improve images with unacceptable level

of noise.

Materials and Methods:

Three clinical protocols are used on Optima

CT540 system: axial Routine Head, Helical Head, Abdomen Pelvis. DCT10

and CTSD16 RTI probes in PMMA phantoms for head and body CTDI

measurements are used for each clinical protocol at the available dose

reduction levels (0, 20, 30, 40%). The measured CTDI are compared with

the expected reduction. A Cathpan 600 phantom is scanned at the

above levels and the analysis is performed with AutoQA Lite™-IrisQA

software. Furthermore, ASiR is tested on image data obtained at 0% dose

reduction level, applying retrospectively all the available steps of noise

reduction.

Results:

The measured CTDI confirm the nominal dose reduction levels and

only at the upper dose reduction, noise level shows an increase of 7% in

the axial Routine Head protocol, while for all clinical protocols uniformi-

ty and low contrast detectability (LCD) are unchanged and MTF50 improves

up to 20%. Referring to quality of reconstructed images, depending on the

steps of noise reduction, all parameters are improved: MTF50 grows up to

23% in Abdomen Pelvis protocol, noise is reduced to 33% in Helical Head

protocol, LCD increases of 43% in Abdomen Pelvis protocol, while unifor-

mity is always unchanged.

Conclusion:

CT extensive use in medical procedures contributes about 50%

of the population’s medical exposure. The raising radiation dose to the

patient population from CT has created a major concern in the medical

imaging community in an effort to optimize procedures and to reduce ra-

diation risk.

e87

Abstracts/Physica Medica 32 (2016) e71–e96