Table of Contents Table of Contents
Previous Page  88 / 146 Next Page
Information
Show Menu
Previous Page 88 / 146 Next Page
Page Background

Conclusions:

The DMS has supported the implementation of iterative re-

construction system, to reduce CT dose. A systematic control of patient dose,

using dose monitoring software in Radiology Service, supports the process

of optimization of radiological practice.

http://dx.doi.org/10.1016/j.ejmp.2016.01.283

B.279

EFFECT OF THE PEDIATRIC PATIENT POSITIONING AND MANAGEMENT

ON CT AEC

A. Mari

* , a ,

E. Cerquetell

a b ,

S. Maggi

a . a

A.O.U. Ospedali Riuniti, Ancona, Italy;

b

Universita’ Politecnica Delle Marche, Ancona, Italy

Introduction:

Because of the high radiosensitivity in pediatrics it is nec-

essary to optimize CT exams and know which would be the impact of the

patient positioning on the AEC.

Materials and Methods:

We evaluated the dose on a GE Lightspeed 16 slices

on a pediatric anthropomorphic phantom (10 years old, AARON model 705-

D, Atom Inc.) varying the following parameters: 80–100 kV,10–15–20 NI,

10–20 cm collimation. We changed patient offset (

±

5 cm AP and LR) and

scout order. For all the acquisitions SmartAm and AutoAm (GE’s AEC) were

used. The dose was calculated from mAs/slice multiplied by a factor from

QC. We calculated the % difference in CTDI between correct and mis-

aligned position.

Results:

For vertical misalignment at 100 kV (AP-LL scout) analysis shows

a mean CTDI variation of 10% without a clear trend, more variations are

present on morphological and density discontinuity (40%), for higher NI

and small collimation that trend increase of less than 5%. When scout is

LL-AP the trend is more clear and scans with patient positioned

+

5 cm have

of dose of meanly

+

15%, patient placed

5 cm have CDTI/slice

20% respect

patient at isocenter. Dose delivery is more stable for low NI and small col-

limation. At 80 kV dose delivery follows the same trends of 100 kV scans

with 5% of more difference (in each direction) and more variability. With

lateral misalignment dose varies (until 20%) for LL/AP scans but without

difference between the two positions, when the second scout is the lateral

the difference is about

±

10% but less stable, inverting scouts the differ-

ence between the two position is evident but not always higher or lower

respect the isocentric position, values variations reach the 30%.

Conclusions:

Alignment and scout order may change significantly dose de-

livery for the same exam on the same patient. Protocols must be carefully

applied and configured, when is not possible to place correctly the patient

that variations must be taken in account and corrected by protocol set up.

http://dx.doi.org/10.1016/j.ejmp.2016.01.284

B.280

PATIENT CT DOSIMETRY: USE OF THE DICOMSR PROTOCOL FOR

CALCULATION AND MONITORING OF DRL AND RELEVANT PARAMETERS

A. Martignano

*

, A. Valentini.

Department of Medical Physics, Azienda

Provinciale per i Servizi Sanitari, APSS, Trento, Italy

Introduction:

The Dicom Structured Report (DicomSR) is a format capable

of carrying several types of information. Of particular interest is the pos-

sibility to associate structured dosimetric information to each diagnostic

exam. The data from the CT scanners of the APSS capable of sending DicomSr,

are collected and analyzed in order to obtain useful information about the

use of the scanners and the dose received by the patients.

Materials and Methods:

An Excel software was written to organize and

analyze the data. The program recollects all extracted data, separates them

on the basis of several parameters and performs statistical and dosimet-

ric analysis. Annual reports are written. The reports show statistics on the

number of patients and dose events, the scanned anatomic region, the

number of scans per exams for different anatomic regions, the average

CTDIvol and the DRL of the DLP value. The Excel software calculates the

mean effective dose on the basis of EC 2000 dose model. DLP values are

compared to DRL limits.

Results:

41,712 exams and 144,252 dose events were analyzed in 3 years

on 5 CT. The mean effective dose per dose event was 2.1 mSv and the mean

effective dose per patient was 7.2 mSv. An incorrect protocol without current

optimization was identified and corrected on the basis of a difference in

the DLP value with respect to the other scanners. DLP DRL values per dose

event were lower than the limit values for all types of exams.

Conclusions:

The DicomSr format showed to be a very useful tool to re-

trieve and analyze dosimetric information, giving the possibility to perform

analysis and identify issues. However a software is needed to manage the

data. An Excel software was written to perform statistical analysis and cal-

culate effective doses to the patients. Annual reports with the main

information are written and sent to the involved professionals in order to

optimize the diagnostic process.

http://dx.doi.org/10.1016/j.ejmp.2016.01.285

B.281

THE SYRMA-CT PROJECT FOR BREAST COMPUTED TOMOGRAPHY WITH

SYNCHROTRON RADIATION: IMAGE QUALITY AND DOSE EVALUATION

G. Mettivier

*

.

Università degli Studi di Napoli Federico II, Napoli, Italy

Introduction:

The SYRMA-CT project (INFN, ELETTRA Synchrotron Radi-

ation Facility, Trieste University Hospital) is a pilot study for 3D breast

imaging with Synchrotron Radiation (SR) monochromatic X-rays, via prop-

agation based phase-contrast X-ray CT of the pendant breast. This project

aims at producing the first high-resolution phase-contrast in vivo images

of the breast, for improved visualization of breast cancer lesions, at dose

levels comparable to that of two-view mammography.

Materials and Methods:

In SR breast CT the patient is in prone position

with her breast freely pending without compression from a hole in the bed.

The SR beam is incident horizontally and irradiates a thin section of the

breast. For acquiring multiple views, the patient bed rotates around a hor-

izontal axis: for a scan of a thick section of the breast the bed translates

vertically. The SR laminar beam has a size of 210

×

3 mm

2

and the photon

energy was 38 keV. The detector is at about 2 m away from the isocenter.

The single-photon counting detector is PIXIRAD-8, a 0.65-mm thick CdTe

hybrid detector (25

×

2.5 cm

2

sensitive area) with 60-μm pixel pitch on a

hexagonal 2D arrangement. A breast specimen of about 9 cm diameter was

imaged, as well as test objects for image quality evaluation. CT slices were

reconstructed via FBP or iteratively both in attenuation and after phase re-

trieval using the homogeneous transport-of-intensity algorithm.

Discussion and Conclusion:

First CT attenuation images and phase images

have been produced of the breast specimen, at glandular dose levels in the

irradiated slice of 5–20 mGy. Phase retrieval increased the signal-to-

noise ratio with respect to attenuation CT; iterative reconstruction allowed

to reduce significantly the dose with a limited reduction of the contrast-

to-noise ratio. The limiting spatial resolution was over 2.5 mm-1 after phase

retrieval. SR phase-contrast breast CT is technically feasible at dose levels

comparable to two-view mammography.

http://dx.doi.org/10.1016/j.ejmp.2016.01.286

B.282

CT-BASED MONITORING OF MICROWAVE THERMAL ABLATION

S. Minosse

* , a ,

D. D’Alessi

o b , V

. Loprest

o c , R

. Pinto

c , L

. Farina

d , M

. Cavagnaro

d ,

L. Strigari

b .

a

Post Graduate School of Medical Physics, Sapienza University of

Rome, Rome, Italy;

b

Medical Physics Laboratory, Regina Elena National Cancer

Institute, Rome, Italy;

c

ENEA, Division of Health Protection Technologies,

Casaccia Research Center, Rome, Italy;

d

Department of Information Engineering,

Electronics and Telecommunications, Sapienza University of Rome, Rome, Italy

Introduction:

Microwave thermal ablation (MWA) is a therapeutic tech-

nique which exploits the absorption of electromagnetic energy at microwave

frequencies to induce an irreversible coagulation necrosis in the target tissue

through a temperature increase higher than 55–60°C. Recent research has

been focused on increasing the efficiency of MWA procedures and reduc-

ing the tumors recurrence rate through a full control of the thermally ablated

region. The combination of CT imaging and MWA is used to examine spatial

evolution of the thermally ablated volume during treatment.

Materials and Methods:

A plastic box with a regular grid of holes on its

walls was designed to contain tissue samples and to insert lead fiducial

markers and an antenna applicator. A CT system was used for real-time

monitoring of density variation during the ablation procedure and of the

movements of the lead markers. A software was developed in LabView, syn-

chronized with the CT acquisition, for automatic remote control of the MWA

e83

Abstracts/Physica Medica 32 (2016) e71–e96