Table of Contents Table of Contents
Previous Page  73 / 146 Next Page
Information
Show Menu
Previous Page 73 / 146 Next Page
Page Background

Results:

Average percentage volumes covered by the prescribed dose

were significantly different in RT and MCrecalc plans for different tumor

size for GTV (GTVsmall_V

= −

17.4

±

18.8%, GTVlarge_V

= −

6.6

±

7.4%, p

<

0.05)

and PTV (PTVsmall_V

= −

51.9

±

21.9%, PTVlarge_V

= −

23.7

±

12.1%, p

<

0.01).

For GTV, average Dmean and D50 differed between RT and MCrecalc

plans according to tumor size (small: Dmean

= −

11.8

±

6.7%,

D50

= −

11.5

±

6.6%; large: Dmean

= −

5.1

±

2.1%, D50

= −

5.0

±

2.1%, p

<

0.01).

Analogous behavior was identified for PTV (small: Dmean

= −

18.6

±

9.6%,

D50

= −

18.5

±

10.7; large: Dmean

= −

7.0

±

2.9%, D50

= −

6.3

±

2.9%, p

<

0.01).

Differences for lungs in V20, V10, and V5 are

0.6

±

0.4%,

1.4

±

1.4% and

3.4

±

4.4%, respectively.

Conclusion:

Both GTV Dmean and D50 could be used to evaluate target

dose underestimation for lung lesions. Based on GTV dose parameters dif-

ference between RT and MC calculations, a different correction to dose

prescription for small and large tumors could be adopted. The relation to

tumor position and beam angles geometry should be further investi-

gated. Target under-coverage suggests that optimization with MC algorithm

needs to be introduced in treatment planning.

http://dx.doi.org/10.1016/j.ejmp.2016.01.232

A.229

A VERSATILE ASYMMETRIC COLLIMATION SYSTEM FOR CLINICAL USE IN

EXTERNAL ELECTRON BEAM RADIOTHERAPY WITH ELEKTA’S LINAC

A. Valentini

* , a ,

L. Menegotti

a ,

D. Ravanelli

a ,

K. Najera

b , c .

a

Medical Physics

Department, Azienda Provinciale per i Servizi Sanitari – APSS, Trento, Italy;

b

The Abdus Salam International Centre for Theoretical Physics – ICTP, Trieste,

Italy;

c

Radiotherapy Department, Instituto de Cancerologia – Dr. Bernardo del

Valle S., Guatemala, CA, Guatemala

Introduction:

In June 1996 the Medical Physics Department (MPD) of S.

Chiara Hospital in Trento designed and realized an electron collimator to

be adapted on a standard Elekta’s Linac cone applicator for electron beam

radiotherapy treatments. The major innovation was the variable asym-

metric rectangular collimation of the electron beam, not offered by the

vendors until today. Another outstanding feature is its interchangeability

between the two different Elekta’s Linacs installed at S. Chiara Hospital in

Trento used for electron therapy.

Materials and methods:

Electron beams were used firstly for patient spe-

cific head and neck boost and now mostly for patient specific breast boost.

For such treatments the MPD had the clinical necessity to create an asym-

metric collimator and designed it since no vendors offered such products.

Brass knurled was used to build a variable asymmetrical collimator, to be

fixed to the standard squared electron applicator of 10 cm

×

10 cm.

Results:

The collimation of electron beams is obtained by moving two or-

thogonal knurled brass plates of 8.2 mm thickness, resulting in 50 different

field collimation shapes and sizes within squared electron applicator of

10 cm

×

10 cm. From June 1996 up to now, approximately 6000 electron

treatments were delivered using this collimator, i.e. 80% of the electron treat-

ments performed at the Radiotherapy Department at S. Chiara Hospital in

Trento. The replacement of lead blocks shielding with the new asymmet-

ric collimation system for electron beams, leading to spare time and money,

improving the reproducibility.

Conclusions:

Nowadays this electron collimator is still used in clinical elec-

tron treatments due to its ease of use and practicality. For 19 years there

were no problems related to its clinical use, showing high reliability and

mechanical robustness.

http://dx.doi.org/10.1016/j.ejmp.2016.01.233

A.230

COMMISSIONING OF FLATTENING FILTER FREE BEAM FOR STEREOTACTIC

ABLATIVE RADIOTHERAPY (SABR) LUNG TREATMENTS

S. Valzano

*

, a ,

M. Fusella

b ,

E. Mones

b ,

C. Secco

b ,

F. Puricelli

b ,

G. Loi

b ,

M. Brambilla

b .

a

Medical Physics Unit, AO Ordine Mauriziano, Turin, Italy;

b

Medical Physics Unit, AOU Maggiore della Carità, Novara, Italy

Introduction:

To commission a 6 MV flattening filter free (FFF) photon beam

from a LINAC Varian Trilogy TX in the RayStation TPS.

Materials:

In order to commission the FFF beam in a TPS, based on a three

source model, PDD, profiles and output factos (OF) were acquired in a water

phantom with isocentric setup. The following detectors were used: an IBA

CC13 chamber, an Exradin A16 μchamber, a Sun Nuclear edge diode (SE),

an IBA diode (SFD) and an Exradin W1 scintillator (ES). CC13 was used in

large field dosimetry, for small field sizes, ranging from 0.6 cm

2

to 3 cm

2

;

all the other detectors were used and their outcomes compared in order

to study and minimize the impact of dose uncertainties on the beammodel.

The commissioning was validated by end to end tests in simple and complex

geometry in phantom. VMAT SABRT was planned using the FFF beam in

phantom and real patients and compared with the standard (FF) beam plans

by means of DVH analysis and patient specific QA passing rate obtained

with ArcCHECK.

Results:

The SE diode performed better than A16 and SFD in small fields

with reduced partial volume effects and less energy dependence, then it

was our detector of choice for PDD and profile scans. The diode OFs were

in agreement within 2% with ES down to the field 0.8 cm

2

. Only for the

smallest field the ED and SFD overestimate the ES by 11% and 7%

respectively; the ED reading was corrected on the basis of the other two

ones. The beam commissioning using these data met successfully the

Van Dyk criteria with minimal OF corrections between calculated and

measured data. End to end test showed good agreement between calcu-

lated and measured doses with passing rates

>

90% for gamma(2/2)

criteria. No significant differences were found between FFF and FF plans

QA passing rates and in DVH analysis.

Conclusions:

FFF beams can be modeled in RayStation with dosimetric per-

formances comparable with FF beams in lung SABR, but with relevant time

sparing and improved treatment accuracy.

http://dx.doi.org/10.1016/j.ejmp.2016.01.234

A.231

STUDY OF THE DOSE DELIVERY SYSTEM INACCURACIES AND THEIR

IMPACT ON THE DOSE DISTRIBUTION DURING THE FIRST YEARS OF THE

CNAO CLINICAL ACTIVITY

A. Vignati

* , a ,

M.A. Hosseini

b ,

A. Attili

a ,

M. Donetti

c ,

S. Giordanengo

a ,

L.F. Guarachi

d ,

F. Marchett

o a ,

F. Mas Milian

e ,

A. Mirandola

c ,

S. Molinelli

c ,

V. Monac

o a , d ,

G. Russo

f ,

R. Sacch

i a , d ,

M. Varasteh Anva

r d ,

R. Cirio

a , d .

a

Istituto

Nazionale di Fisica Nucleare, Torino, Italy;

b

Ionizing and Non-ionizing Radiation

Protection Research Center (INIRPRC), Shiraz University of Medical Sciences

(SUMS), Shiraz, Iran;

c

Centro Nazionale di Adroterapia Oncologica (CNAO),

Pavia, Italy;

d

Università degli studi di Torino, Torino, Italy;

e

CNPq Fellow,

Universidade Estadual de Santa Cruz, Bahia, Brazil;

f

Internet-Simulation

Evaluation Envision (I-SEE) s.r.l., Torino, Italy

Introduction:

The Italian hadrontherapy center (CNAO) uses actively scanned

proton and carbon-ion beams to treat tumors and is equipped with a dose

delivery system (DDS) to monitor and guide the beams to the patient. This

work aims at evaluating the impact of the DDS inaccuracies on the dose

distribution of patients treated at CNAO through a retrospective analysis

of the data collected during the delivery (D) and their comparison with the

planning (P).

Materials and methods:

The DDS delivers the dose in spots according to

the P, each spot being defined by the number of particles, the beam posi-

tion and energy. The same quantities are measured by the DDS during the

D and saved in the LOG files. The absolute difference between D- and P-spot

positions at the reference plane at the isocenter, and the relative differ-

ence between the P- and D-number of particles were assessed. Then, the

P- and D-spot quantities were used as inputs of a Forward Planning program,

which computes the dose of a treatment configuration, based on the patient

anatomy and the beams set-up. The obtained P- and D-dose maps were

compared through the Gamma-index, which measures the discrepancy

between two dose distributions, in terms of spatial distance and dose

difference.

Results:

61 patients (01/2012–04/2013) were studied. 98% of the spots

showed position deviations less than 1.5 mm, and differences in number

of particles were less than 2.5%. The P- and D-dose maps were compared

for 5 patients, representing time periods with different accuracy level of

the DDS. Even in the worst case, more than 98% of points passed the (3 mm,

3%) Gamma-index criteria.

Conclusions:

The accuracy of the CNAO DDS was evaluated in terms of de-

viation of the spot position and number of particles between P- and D-spots.

Variations in the position accuracy were correlated with fine tunings of the

e68

Abstracts/Physica Medica 32 (2016) e1–e70