Table of Contents Table of Contents
Previous Page  31 / 146 Next Page
Information
Show Menu
Previous Page 31 / 146 Next Page
Page Background

out to be more robust mainly for more complex MLC shape. The results

showed that both tools are able to successfully evaluate leaves position and

collimator rotation angle with accuracy less than 1 mm and 1 degree re-

spectively. In error free pattern, the estimated difference between planned

and measured MLC position was within

±

1 mm.

Conclusion:

A robust and efficient EPID image analysis tool was devel-

oped to automatically extract an image contour and MLC position. Depending

on the image acquisition device resolution a different edge method should

be applied. The results confirmed that this code is suitable and can be simply

implemented in a QA machine program.

The same procedure could be further developed to check MLC leaves po-

sition during VMAT delivery.

http://dx.doi.org/10.1016/j.ejmp.2016.01.089

A.86

IMPACT OF AUTO BEAM-OFF AND 4D MODEL AUTOMATIC UPDATE ON

TRACKING ACCURACY OF THE VERO SYSTEM

C. Garibaldi

* , a ,

A. Bazani

a ,

F. Pansin

i a ,

R. Ricotti

a ,

D. Ciardo

a ,

S. Com

i a ,

G. Piperno

a ,

A.M. Ferrari

a ,

M. Cremonesi

a ,

B.A. Jereczek-Fossa

a , b ,

R. Orecchia

a , b .

a

Istituto Europeo di Oncologia, Milano, Italy;

b

Università degli

studi di Milano, Milano, Italy

Purpose:

To evaluate the impact of auto beam-off and automatic update

of the 4D model during treatment on the accuracy of dynamic tumor track-

ing (DTT) of the VERO system.

Material and Methods:

We evaluated the tracking and prediction errors

(during the 4D model building and treatment (ET, EP) analysing the syn-

chronized log files. We simulated significant variations of the breathing

pattern: baseline drift, change of amplitude and period and introduction

of a phase shift between internal and external movement. We evaluated

ET, considering 3 scenarios: (a) no auto beam-off and no model update;

(b) auto beam-off and no model update; and (c) auto beam-off and model

update. We evaluated the influence of the number of samples after which

the beam is turned off if outside the threshold (2 mm) on ET and irradia-

tion time and how many updates were necessary to re-establish a good

correlation model. The impact of the monitoring frequency (1 or 2 Hz) on

the accuracy of the update of the model was also evaluated. Tests were per-

formed so far with sinusoidal patterns, but patient’s respiratory patterns

will be evaluated to determine the best parameters for the clinical setting.

Results:

A total of 171 log files were analysed. The mean values of

EP,4Dmodel, EP,treat, and ET were 0.7

±

0.5 mm, 0.6

±

0.4 mm, 0.6

±

0.4 mm,

respectively. The auto beam-off does not seem to reduce ET considerably

in case of a large variation of the breathing pattern even reducing the

number of samples from 3 to 1, while the corresponding treatment times

increased significantly (35 vs. 123 s). According to the type of change, up

to 4 updates are necessary to restore a good correlation model before the

treatment can be restarted. Increasing the monitoring frequency from 1 to

2 Hz does not seem to decrease ET when using auto beam off.

Conclusions:

The automatic update of the 4D model is a powerful tool to

guarantee the accuracy of DTT without increasing the imaging dose due

to fluoroscopy used to build new 4D models.

http://dx.doi.org/10.1016/j.ejmp.2016.01.090

A.87

COMPARISON OF FIELD-IN-FIELD TANGENTIAL TREATMENT VERSUS THE

CONVENTIONAL TREATMENT

D. Gaudino

*

, L. Bellesi, G. Stimato, C. Di Venanzio, A. Mameli, E. Infusino,

E. Ippolito, S. Silipigni, C. Rinaldi, S. Ramella, L. Trodella,

R.M. D’Angelillo.

Universita’ Campus Biomedico Di Roma, Roma, Italy

Purpose/Objective:

Field-in-field (FIF) technique ameliorates convention-

al planning with tangential fields (TANG) for adjuvant treatments of breast

cancers. It consists of the application of additional fields in order to improve

dosimetric parameters. Either FIF or TANG is evaluated comparing dose dis-

tributions on PTV, OAR and DVH constraints. We evaluated retrospectively

the statistical significance of such differences.

Material and methods:

33 patients were evaluated. Endpoints evaluated

were: V95, V105, maximum dose within PTV, maximum dose, lung

maximum dose, lung mean dose, heart maximum dose, and heart mean

dose. Geometrical misalignment was evaluated by EPI. The baseline FIF was

compared to the TANG plan. FIF was recalculated on TPS incorporating the

misalignment data (FIFErrors). A statistical analysis comparing TANG and

FIFErrors results was addressed by Wilcoxon.

Results:

We analyzed misalignment data in 33 patients. Mean values for

FIF and TANG plans were respectively: V95

=

98.92 versus 98.25%; maximum

dose

=

109.0 versus 110.01%; maximum dose within PTV

=

108.32 versus

109.01; V105

=

4.01 versus 4.42. The FIF was significantly superior to the

TANG plan for V95 (p

=

0.003), maximum dose (p

=

0.002), maximum dose

within PTV (p

=

0.033); it was not significantly superior for V105 (p

=

0.201),

although the mean V105 value was overall inferior for the FIF (4.01% FIF

versus 4.42% TANG).

Themean gainby the adoptionof FIF over the TANGaccounted for V95

=

0.67%;

maximum dose

=

1.01%; maximum dose within PTV

=

0.69%; V105

=

0.41%.

Once recalculated considering themisalignment it was reduced by 2.98% for

V95, 10.14% for maximum dose; 7.93% for maximum dose within PTV; and

24.39% for V105, respectively. SOFTDISO softwarewas used in order to control

correct positioning of patients day by day. Results were analyzed.

Conclusion:

FIF technique optimizes the planning and presents a good ge-

ometrical stability, while the impact on organs at risk requires further

evaluation.

http://dx.doi.org/10.1016/j.ejmp.2016.01.091

A.88

IMPACT OF THE MLC DELIVERY ERRORS ON PATIENT DOSE FOR IMRT

TREATMENTS: A COMPARISON BETWEEN PLANNED DVH AND

RECONSTRUCTED DVH BASED ON MLC LOG FILE (DYNALOG)

S. Gelosa

*

, F. Parisoli, P. Lattuada, M. Frigerio, C. Berlusconi,

A. Ostinelli.

Azienda Ospedaliera S. Anna, Como, Italy

Introduction:

MLC position accuracy during sliding window IMRT treat-

ment was verified by the analysis of DynaLog files to compare planned and

recalculated patient dose distributions by DVHs.

Material and methods:

10 clinical plans (breast, pelvis, prostate, lymph-

nodes) were reconstructed to compare the dose distribution of the reference

plan (clinical plan) with the recalculated one by TPS Eclipse (Varian). Plans

were delivered by Varian Clinac iX and the DynaLog files generated by the

controller were imported in LINACwatch software (QualiFormeD, vers.1.2)

to create a RT plan for each session. These were recalculated on patient

CT with the same dose calculation algorithm (AAA13.0.26) and monitor units

as the reference plan.

Results:

PTV DVH percentage variations are less than 1% for D95%, D98%,

D2% and Dm, except for breast treatment involving lymph-nodes where

the maximum variation is 2% for lymph-node PTV. The 90–120% isodose

lines of reconstructed plans are wider, improving target coverage without

substantial variations in OAR DVH.

Reconstructed DVH differences between sessions were negligible, with one

exception where little variation in OAR was observed when the delivery

was interrupted.

Conclusion:

In the optimization of breast involving lymph-nodes, the OAR

constraint observance may produce MLC movements hard to reproduce,

explaining the difference in PTV lymph-nodes. If the optimization is not

stressed, DVH variation is nonsignificant. This situation is similar for pelvis

when the objective on bowels is difficult to reach. This software allows to

perform patient specific IMRT QA by comparing DVHs, but it does not replace

pre-treatment dosimetric verifications.

http://dx.doi.org/10.1016/j.ejmp.2016.01.092

A.89

VERIFICATION OF DOSE DISTRIBUTION FROM CCX RU-106 EYE-PLAQUES

BY USING A MICRODIAMOND DOSIMETER

E. Genovese

* , a ,

M. Pimpinella

b ,

A.S. Guerra

b ,

V. De Coste

b ,

M. Marinelli

c ,

G. Verona Rinat

i c ,

S. Donatiello

a ,

C. Orlandi

a ,

A. Romanzo

d ,

R. Cozz

a d ,

V. Cannata

a .

a

Enterprise Risk Management/Medical Physics, Bambino Gesù

Children’s Hospital, IRCCS, Rome, Italy;

b

ENEA, National Institute of Ionizing

Radiation Metrology (INMRI) Casaccia, Rome, Italy;

c

Department of Industrial

Engineering, University of Rome Tor Vergata, Rome, Italy;

d

Bambino Gesù

Children’s Hospital, IRCCS, Rome, Italy

e26

Abstracts/Physica Medica 32 (2016) e1–e70