Result:
The dosimetric comparison supplied agreements within 2% for the
isocenter doses; gamma % greater than 0.97% for head and neck, 98% for
lung and 99% for pelvic regions and the gamma-mean values within 0.4.
The PTV V95 and mean doses were within 2%, while the mean doses of
principal OARs were within 3%.
Conclusion:
The CBCT calibration method is accurate enough to supply
useful information for adaptive radiotherapy strategy.
http://dx.doi.org/10.1016/j.ejmp.2016.01.079A.76
IMPROVING THE QUALITY OF PROTONTHERAPY TREATMENT PLANS AND
THEIR VERIFICATION WITH MONTE CARLO METHODS
F. Fracchiolla
* , a , b ,M. Schwarz
a , c .a
Azienda Provinciale per i Servizi Sanitari
(APSS), Trento, Italy;
b
Post Graduate School of Medical Physics ‘Università La
Sapienza’, Rome, Italy;
c
INFN-TIFPA, Trento, Italy
Introduction:
In radiotherapy an independent tool for MU verification is
necessary to ensure the best treatment safety. We developed
[1]a Monte
Carlo (MC) tool for independent dose calculation and we propose here
further applications to the clinical practice of our protontherapy center.
Methods andmaterials:
We used the MC code for three clinical applications:
•
Recalculation of a nominal plan from the TPS;
•
Recalculation of a treatment session by using log file from the beam
delivery system;
•
Simulation of patient specific QA (PQA) procedure (retrospective study
on 27 fields for 10 patients). For each PQA TPS and MC dose distribu-
tions calculated in water equivalent material at different depth were
compared with measurements performed with an array of 1020 ion-
ization chamber detector via γ analysis (3%, 3 mm). In our PQA protocol
the γ passing rate (PR) has decision thresholds:
•
PR
>
95%: field accepted;
•
90%
<
PR
<
95%: if accepted, a justification must be added to PQA report;
•
PR
<
90%: field rejected.Measurement and simulation times were com-
pared too.
Results:
For a neuroma case we recalculated the nominal plan and the dose
based on the log file. The recalculation based on the log file showed a re-
duction of max dose (1 Gy) in brainstem and in the fifth cranial nerve.
MC has a PR always greater than 95% for every depth, while TPS results
are always in the range of 90–95%. The measurement time of a PQA takes
almost 1.5 hours, while the simulations can be performed in parallel and
take 15 min on average.
Conclusion:
Our tool let us to estimate the effects on the dose distribu-
tion due to delivery fluctuations of the machine.
We also proposed a method to drastically reduce PQA verification time. Our
suggestion is to substitute measurements with simulations that showed
a very high accordance in terms of γ PR; only one field per patient should
be measured at one single depth as safety check.
Reference
[1]
Fracchiolla F, Lorentini S, Widesott L, Schwarz M. Characterization and validation of a Monte Carlo code for independent dose calculation in proton therapy treatments with pencil beam scanning. Phys Med Biol 2015;60:8601–19. http://dx.doi.org/10.1016/j.ejmp.2016.01.080A.77
A NEW PHANTOM FOR DAILY QA IN PROTONTHERAPY: A FAST, RELIABLE
AND INEXPENSIVE SOLUTION
F. Fracchiolla
* , a , b ,N. Bizzocch
i a ,C. Algranat
i a ,M. Schwarz
a , c .a
Protontherapy
Department, Azienda Provinciale per i Servizi Sanitari (APSS), Trento, Italy;
b
Post
Graduate School of Medical Physics, ‘Università La Sapienza’, Roma, Italy;
c
INFN-TIFPA, Via Sommarive 14, 38123 Trento, Italy
Introduction:
In a radiotherapy center daily QA (DQA) measurements need
to be fast, accurate and sensitive to any variation of the output of the de-
livery machine that may have a clinical impact. In a pencil beam scanning
protontherapy center, spot positioning, spot size, range and dose output
are usually verified every day before treatments. We designed and built a
new reliable, sensitive and inexpensive phantom for DQAs that reduces the
execution times preserving the accuracy of the test.
Materials and methods:
The phantom is provided with two couple of
wedges: thanks to these the Bragg peak is sampled at different depths and
the image becomes a transposition on the transverse plane of the depth
dose. Three ‘boxes’ in the center are used to check spot positioning and de-
livered dose. The box thickness helps ‘spreading’ the single spot and to fit
a Gaussian profile on a low resolution (7.6 mm) 2D detector.
We tested whether our new QA solution could detect errors of 1 mm in
spot positioning, 2 mm in range and 10% in spot size. Execution time was
also investigated and compared with the previous workflow.
Results:
Our method is able to correctly detect 98% of spots that are ac-
tually in tolerance and 99% of spots out of 1 mm tolerance. All range
variations greater than the threshold (2 mm) were correctly detected. A
month of DQA showed that the repeatability of the sigma is very high (the
standard deviation of each spot is less than 0.7%). The aim of the sigma
analysis is to check the constancy of this parameter: it is not a check of
the sigma in air but in a medium. The sigma in air is measured during weekly
QA.
This phantom let us halve the execution time (20 min on average) com-
pared with the former multi-device procedure.
Conclusions:
We designed and built a phantom precise, accurate, inex-
pensive and fast in terms of execution time; thanks to our new procedure,
we can preserve the reliability of DQAs and save time, favoring more slots
for patient treatments.
http://dx.doi.org/10.1016/j.ejmp.2016.01.081A.78
PATIENT QUALITY ASSURANCE FOR HD TOMOTHERAPY®: A 3-YEAR
REVIEW OF PRE-TREATMENT IN-PHANTOM DOSIMETRY
C. Fulcheri
*
, A. Chiappiniello, M. Marcantonini, C. Zucchetti, M. Iacco,
A.C. Dipilato, A. Didona, R. Tarducci.
Medical Physics Unit, Santa Maria della
Misericordia Hospital, Perugia, Italy
Introduction:
The purpose of this study was to analyze the results of de-
livery quality assurance (DQA) dosimetry performed at our hospital over
the first 3 years of TomoTherapy® use.
Materials and methods:
349 Tomotherapy® DQA results were consid-
ered. Patient DQAs were grouped according to pathology and irradiated site,
which implies taking into account the DQA dependences on the machine
parameters, the dose prescription and the in-axis/off-axis position of ir-
radiated volumes. The DQA measurements were performed with an
ArcCHECK dosimetry phantom in combination with an internal A1SL ion-
ization chamber. Comparisons between planned and measured dose
distributions were performed using a gamma index metric of 3%/3 mm in
relative dose only, with a 10% low dose threshold. For absolute dose com-
parisons, the relative differences (RDs) between measured and calculated
doses were assessed. Statistical analysis was performed with the non-
parametric Friedman test.
Results:
All the plans analyzed passed our quality criteria: gamma passing
rate GP
>
95% and RD
<
3%. A statistically significant difference in DQA results
was found among the different sites: the median values of GP resulted 99.5%
and 99.4% for the breast and thoracic wall (with or without lymph nodes),
respectively, while for all other sites it resulted to being always greater than
99.8%. Likewise, similar results were observed for RD: 1.7% for the breast
and 1.9% for the thoracic wall, while for all other sites it was always
<
1%.
The authors hypothesize that the lower agreement in dose and dose dis-
tributions recorded for the breast and thoracic wall may be ascribed to the
off-axis positions of the lesions. Further investigations are required in this
issue.
Conclusions:
In this review patient DQAs were always found to satisfy our
quality criteria; however, at the same time differences in RD and GP values
among different sites were observed. These results could be used to rede-
sign DQA frequencies and/or tolerances.
http://dx.doi.org/10.1016/j.ejmp.2016.01.082e23
Abstracts/Physica Medica 32 (2016) e1–e70




