Table of Contents Table of Contents
Previous Page  28 / 146 Next Page
Information
Show Menu
Previous Page 28 / 146 Next Page
Page Background

Result:

The dosimetric comparison supplied agreements within 2% for the

isocenter doses; gamma % greater than 0.97% for head and neck, 98% for

lung and 99% for pelvic regions and the gamma-mean values within 0.4.

The PTV V95 and mean doses were within 2%, while the mean doses of

principal OARs were within 3%.

Conclusion:

The CBCT calibration method is accurate enough to supply

useful information for adaptive radiotherapy strategy.

http://dx.doi.org/10.1016/j.ejmp.2016.01.079

A.76

IMPROVING THE QUALITY OF PROTONTHERAPY TREATMENT PLANS AND

THEIR VERIFICATION WITH MONTE CARLO METHODS

F. Fracchiolla

* , a , b ,

M. Schwarz

a , c .

a

Azienda Provinciale per i Servizi Sanitari

(APSS), Trento, Italy;

b

Post Graduate School of Medical Physics ‘Università La

Sapienza’, Rome, Italy;

c

INFN-TIFPA, Trento, Italy

Introduction:

In radiotherapy an independent tool for MU verification is

necessary to ensure the best treatment safety. We developed

[1]

a Monte

Carlo (MC) tool for independent dose calculation and we propose here

further applications to the clinical practice of our protontherapy center.

Methods andmaterials:

We used the MC code for three clinical applications:

Recalculation of a nominal plan from the TPS;

Recalculation of a treatment session by using log file from the beam

delivery system;

Simulation of patient specific QA (PQA) procedure (retrospective study

on 27 fields for 10 patients). For each PQA TPS and MC dose distribu-

tions calculated in water equivalent material at different depth were

compared with measurements performed with an array of 1020 ion-

ization chamber detector via γ analysis (3%, 3 mm). In our PQA protocol

the γ passing rate (PR) has decision thresholds:

PR

>

95%: field accepted;

90%

<

PR

<

95%: if accepted, a justification must be added to PQA report;

PR

<

90%: field rejected.Measurement and simulation times were com-

pared too.

Results:

For a neuroma case we recalculated the nominal plan and the dose

based on the log file. The recalculation based on the log file showed a re-

duction of max dose (1 Gy) in brainstem and in the fifth cranial nerve.

MC has a PR always greater than 95% for every depth, while TPS results

are always in the range of 90–95%. The measurement time of a PQA takes

almost 1.5 hours, while the simulations can be performed in parallel and

take 15 min on average.

Conclusion:

Our tool let us to estimate the effects on the dose distribu-

tion due to delivery fluctuations of the machine.

We also proposed a method to drastically reduce PQA verification time. Our

suggestion is to substitute measurements with simulations that showed

a very high accordance in terms of γ PR; only one field per patient should

be measured at one single depth as safety check.

Reference

[1]

Fracchiolla F, Lorentini S, Widesott L, Schwarz M. Characterization and validation of a Monte Carlo code for independent dose calculation in proton therapy treatments with pencil beam scanning. Phys Med Biol 2015;60:8601–19. http://dx.doi.org/10.1016/j.ejmp.2016.01.080

A.77

A NEW PHANTOM FOR DAILY QA IN PROTONTHERAPY: A FAST, RELIABLE

AND INEXPENSIVE SOLUTION

F. Fracchiolla

* , a , b ,

N. Bizzocch

i a ,

C. Algranat

i a ,

M. Schwarz

a , c .

a

Protontherapy

Department, Azienda Provinciale per i Servizi Sanitari (APSS), Trento, Italy;

b

Post

Graduate School of Medical Physics, ‘Università La Sapienza’, Roma, Italy;

c

INFN-TIFPA, Via Sommarive 14, 38123 Trento, Italy

Introduction:

In a radiotherapy center daily QA (DQA) measurements need

to be fast, accurate and sensitive to any variation of the output of the de-

livery machine that may have a clinical impact. In a pencil beam scanning

protontherapy center, spot positioning, spot size, range and dose output

are usually verified every day before treatments. We designed and built a

new reliable, sensitive and inexpensive phantom for DQAs that reduces the

execution times preserving the accuracy of the test.

Materials and methods:

The phantom is provided with two couple of

wedges: thanks to these the Bragg peak is sampled at different depths and

the image becomes a transposition on the transverse plane of the depth

dose. Three ‘boxes’ in the center are used to check spot positioning and de-

livered dose. The box thickness helps ‘spreading’ the single spot and to fit

a Gaussian profile on a low resolution (7.6 mm) 2D detector.

We tested whether our new QA solution could detect errors of 1 mm in

spot positioning, 2 mm in range and 10% in spot size. Execution time was

also investigated and compared with the previous workflow.

Results:

Our method is able to correctly detect 98% of spots that are ac-

tually in tolerance and 99% of spots out of 1 mm tolerance. All range

variations greater than the threshold (2 mm) were correctly detected. A

month of DQA showed that the repeatability of the sigma is very high (the

standard deviation of each spot is less than 0.7%). The aim of the sigma

analysis is to check the constancy of this parameter: it is not a check of

the sigma in air but in a medium. The sigma in air is measured during weekly

QA.

This phantom let us halve the execution time (20 min on average) com-

pared with the former multi-device procedure.

Conclusions:

We designed and built a phantom precise, accurate, inex-

pensive and fast in terms of execution time; thanks to our new procedure,

we can preserve the reliability of DQAs and save time, favoring more slots

for patient treatments.

http://dx.doi.org/10.1016/j.ejmp.2016.01.081

A.78

PATIENT QUALITY ASSURANCE FOR HD TOMOTHERAPY®: A 3-YEAR

REVIEW OF PRE-TREATMENT IN-PHANTOM DOSIMETRY

C. Fulcheri

*

, A. Chiappiniello, M. Marcantonini, C. Zucchetti, M. Iacco,

A.C. Dipilato, A. Didona, R. Tarducci.

Medical Physics Unit, Santa Maria della

Misericordia Hospital, Perugia, Italy

Introduction:

The purpose of this study was to analyze the results of de-

livery quality assurance (DQA) dosimetry performed at our hospital over

the first 3 years of TomoTherapy® use.

Materials and methods:

349 Tomotherapy® DQA results were consid-

ered. Patient DQAs were grouped according to pathology and irradiated site,

which implies taking into account the DQA dependences on the machine

parameters, the dose prescription and the in-axis/off-axis position of ir-

radiated volumes. The DQA measurements were performed with an

ArcCHECK dosimetry phantom in combination with an internal A1SL ion-

ization chamber. Comparisons between planned and measured dose

distributions were performed using a gamma index metric of 3%/3 mm in

relative dose only, with a 10% low dose threshold. For absolute dose com-

parisons, the relative differences (RDs) between measured and calculated

doses were assessed. Statistical analysis was performed with the non-

parametric Friedman test.

Results:

All the plans analyzed passed our quality criteria: gamma passing

rate GP

>

95% and RD

<

3%. A statistically significant difference in DQA results

was found among the different sites: the median values of GP resulted 99.5%

and 99.4% for the breast and thoracic wall (with or without lymph nodes),

respectively, while for all other sites it resulted to being always greater than

99.8%. Likewise, similar results were observed for RD: 1.7% for the breast

and 1.9% for the thoracic wall, while for all other sites it was always

<

1%.

The authors hypothesize that the lower agreement in dose and dose dis-

tributions recorded for the breast and thoracic wall may be ascribed to the

off-axis positions of the lesions. Further investigations are required in this

issue.

Conclusions:

In this review patient DQAs were always found to satisfy our

quality criteria; however, at the same time differences in RD and GP values

among different sites were observed. These results could be used to rede-

sign DQA frequencies and/or tolerances.

http://dx.doi.org/10.1016/j.ejmp.2016.01.082

e23

Abstracts/Physica Medica 32 (2016) e1–e70