Results:
T2 Reference values were found between 13 and 150 ms. The mean
percentage difference between T2 reference and GRASE values was 16%.
In vivo brain T2 values ranged from 80 to 120 ms. The Bland–Altman plot
showed a positive trend, possibly indicating that the differences from the
two calculation methods tend to get larger as the average increase.
Conclusions:
A number of recent studies reported that brain iron content
can be assessed by T2 measurements. GRASE is a relatively new technique
with the potential to calculate brain volumetric T2 values. However, to the
best of our knowledge, its accuracy has never been tested. We believe that
the accuracy in calculating in vivo T2 values is essential in studies aiming
to assess brain diseases. In our experience, the use of T2 values evaluated
using proprietaryMR software is not recommended in quantitative evaluation.
http://dx.doi.org/10.1016/j.ejmp.2016.01.437E.429
QUALITY ASSURANCE OF PHASED ARRAY COILS: ANALYSIS OF THE NOISE
AMPLIFICATION FACTOR DISTRIBUTION
A. Coniglio
* , a ,G. Vilches Freixas
b , c ,A. Santarelli
a ,M. Ciocca
c ,L. Begnozz
i a .a
Medical Physics Department, S. Giovanni Calibita Fatebenefratelli Hospital,
Roma, Italy;
b
Universitè de Lyon, CREATIS, CNRS, UMR5220, Inserm U1044,
INSA-Lyon, France;
c
Medical Physics Unit, CNAO Foundation, Pavia, Italy
Introduction:
Due to the spatially varying noise across the image and the
noise dependence on the geometry coil inherent to parallel imaging (PI)
methods, conventional signal to noise ratio (SNR) measurements are in-
adequate to test the performance of phased array (PA) coils working with
PI. The aim of this work is to provide a quality assurance protocol for PA
coils working with PI based on the analysis of the noise amplification factor
(g-factor) distribution.
Materials and Methods:
A fast gradient echo sequence was optimized to
calculate pixel-based SNR and g-factor maps using 3 different reduction
factors (R). Acquisitions were performed using both sensitivity encoding
(SENSE) and generalized auto-calibrating partially parallel acquisition
(GRAPPA) algorithms. Over a period of 18 months, data were acquired with
1.5 T and 3 T MR systems using PA head coils with 8 and 32 channels, re-
spectively. A custom Matlab® code was developed to generate SNR and
g-factor maps. A three-parameter (a,b, k) log-logistic distribution was used
to fit the g-factor distribution. Mean baseline values with the correspond-
ing standard deviations were calculated for mean g-value, distribution
parameters and SNR, with the latter evaluated according to the AAPM pro-
tocol. Acquisitions with different coil operation modes were performed.
Results:
Reduction factors of 3 and 4 resulted in more efficient discrimi-
nation between different coil operation modes. In a number of situations,
SNR lay within the baseline interval whereas a and k parameters lay outside:
image reconstruction with 7 of 8 channels (SENSE,R
=
3), acquisition per-
formed with dual-MCM (GRAPPA,R
=
4) and with RSM separate operation
mode (GRAPPA,R
=
3).
Conclusions:
The analysis of the g-factor distribution permitted to distin-
guish different coil operation modes with respect to the conventional SNR
calculation method and mere calculation of the mean g-factor. The pro-
posed approach appeared robust for both MR systems and for two different
PI algorithms.
http://dx.doi.org/10.1016/j.ejmp.2016.01.438E.430
IMPLEMENTING INTRA-OPERATIVE MAGNETIC RESONANCE IMAGING
EQUIPMENT – SAFETY CONCERNS
F. Cretti
* , a ,M. Branchi
a ,G. Manni
a ,S. Canini
a ,F. Biroli
a ,G. Bonaldi
a ,C. Bernucci
a ,M. Passon
i a ,R. Suard
i a ,F. Campanella
b ,P. Colomb
o c .a
Azienda
Ospedaliera Papa Giovanni XXIII, Bergamo, Italy;
b
INAIL Area Ricerca
Certificazione Verifica, Roma, Italy;
c
Azienda Ospedaliera Niguarda Ca’ Granda,
Milano, Italy
Introduction:
Intraoperative MRI represents an evolution of
neuronavigational tools, addressing the goal of accurate targeting of the
region of interest, with the advantage of overcoming problematic discrep-
ancy in the imaged anatomy after opening the cranium. Nevertheless, taking
a MRI scanner to the surgery theater is not a straightforward operation,
due to severe safety constraints. In our institution a
< <
on rail
> >
solution
was adopted, consisting in a moving 1.5 T scanner that can work in two
different positions, in a two-room layout. One position is dedicated to di-
agnostic activity whereas the other one is for intraoperative use exclusively.
The two rooms are separated by a sliding door, kept close during diagnos-
tic activity – in that case the surgery room can work independently – and
open during intraoperative use, when the two rooms are an open space,
with the scanner set in the surgery position. This abstract reports prelim-
inary steps accomplished to get local authority approval.
Material and Methods
: A multidisciplinary team worked to get a project
solution compliant with safety standard. Topics concerned both technical
aspects (aeraulic system, dynamic quench system, Faraday cage continu-
ity, emergency devices, static magnetic field shielding, magnetic
compatibility with the underneath 3T magnet room), and organization
aspects, involving site regulations, staff education, patient management,
surgical devices labeling and administration. Also, INAIL consultation was
availed, with significant contribution to project development.
Results:
A satisfying safety strategy was finally achieved. Also, a specific
surgery procedure was written – as an integrant part of the safety rules –
in order to insure adequate preparation of both patient and surgery area
before scanner positioning. Now the system is being installed.
Conclusion:
Given the complexity of the project, multidisciplinary work
is a prerequisite for successful and safe implementation and operating.
http://dx.doi.org/10.1016/j.ejmp.2016.01.439E.431
TEMPORAL STABILITY QA IN FUNCTIONAL MAGNETIC RESONANCE
IMAGING
L. Fedeli *, S. Busoni, M. Fedi, P. Saletti, A. Taddeucci.
Azienda Ospedaliero
Universitaria Careggi, Firenze, Italy
Introduction:
MRI scanner performance temporal stability in functional
magnetic resonance imaging (fMRI) is a fundamental aspect both in lon-
gitudinal and cross sectional studies, due to the fact that the blood oxygen
level dependent (BOLD) effects are of the order of few percent. A revised
version of Friedman protocol (JMRI, 23–2006) was adopted to check the
stability of scanners on a weekly basis. Results of a two year survey of four
1.5 T MRI clinical scanners are reported.
Materials and Methods:
The purpose of the procedure is to measure the
stability of the scanners in conditions that are as similar as possible to the
clinical settings. The same head coil and scanning sequence used in the fMRI
experiments are employed. A time series of 200 images of the QA vendor
phantom is collected, with an overall acquisition time of 15 minutes. The
automated analysis is performed with a home-made software code and pro-
vides information about signal intensity, uniformity, magnitude spectrum,
radius of decorrelation, SNR and signal fluctuation to noise ratio.
Results:
All four 1.5 T MRI scanner data fluctuations are of the order of a
few per cent, showing a suitable stability over the two year survey. It appears
that there is a correlation between SNR and SFNR (r
>
0.7). The Fourier anal-
ysis does not show periodic noise.
Conclusions:
Regular check of parameter stability provides a strong feed-
back regarding scanner performances, and may allow predicting possible
malfunction. Acquired data may be used in order to assess scanner per-
formance reference levels.
http://dx.doi.org/10.1016/j.ejmp.2016.01.440E.432
SAFETY FOR MRI PATIENTS WITH IMPLANTED MEDICAL DEVICES
E. Genovese
* , a ,A. Napolitano
a ,S. Donatiello
a ,C. Orlandi
a ,P. Toma’
b ,F. Campanella
c ,G. Calcagnin
i d ,F. Censi
d .a
Enterprise Risk Management/
Medical Physics, Bambino Gesù Children’s Hospital, IRCCS, Rome, Italy;
b
Department of Imaging, Bambino Gesù Children’s Hospital, IRCCS, Rome, Italy;
c
Dipartimento Medicina, Epidemiologia, igiene del lavoro ed ambientale, INAIL,
Rome, Italy;
d
Dip. Tecnologie e Salute, Istituto Superiore di Sanità, Rome, Italy
Introduction:
Thanks to the high diagnostic power of magnetic reso-
nance scanners, there is a steady increasing in the number of patients
undergoing magnetic resonance exams while having an implanted medical
device. At the same time, even though a few old regulatory burdens within
the Italian law have been recognized as partially outmoded and misguided
e127
Abstracts/Physica Medica 32 (2016) e124–e134




